Urea Detection in Phosphate Buffer and Artificial Urine: A Simplified Kinetic Model of a pH-Sensitive EISCAP Urea Biosensor
Abstrak
A simplified kinetic model for the quantitative analysis of a potentiometric, pH-based urea biosensor is presented. The device was an electrolyte–insulator–semiconductor capacitor (EISCAP) with a pH-sensitive Ta<sub>2</sub>O<sub>5</sub> gate functionalized by a polyallylamine hydrochloride (PAH)/urease bilayer. Within the steady-state approximation, the kinetic equations yielded an implicit algebraic relation linking the bulk urea concentration to the local pH at the sensor surface. Numerical solution of this equation, combined with a fitting routine, provides the apparent Michaelis–Menten constant (<inline-formula><math xmlns="http://www.w3.org/1998/Math/MathML" display="inline"><semantics><mrow><msub><mi>K</mi><mi>M</mi></msub></mrow></semantics></math></inline-formula>) and the normalized maximum reaction rate (<inline-formula><math xmlns="http://www.w3.org/1998/Math/MathML" display="inline"><semantics><mrow><msub><mover accent="true"><mi>k</mi><mo>¯</mo></mover><mi>V</mi></msub></mrow></semantics></math></inline-formula>). Validation against the literature data confirmed the reliability of the approach. Experimental results were then analyzed in both phosphate buffer (PBS) and artificial urine (AU), covering urea concentrations of 0.1–50 mM. The fitted parameters showed comparable <inline-formula><math xmlns="http://www.w3.org/1998/Math/MathML" display="inline"><semantics><mrow><msub><mi>K</mi><mi>M</mi></msub></mrow></semantics></math></inline-formula> values of 10.9 mM (PBS) and 32.4 mM (AU), but strongly different <inline-formula><math xmlns="http://www.w3.org/1998/Math/MathML" display="inline"><semantics><mrow><msub><mover accent="true"><mi>k</mi><mo>¯</mo></mover><mi>V</mi></msub></mrow></semantics></math></inline-formula> values: <inline-formula><math xmlns="http://www.w3.org/1998/Math/MathML" display="inline"><semantics><mrow><mn>2.2</mn><mo>×</mo><msup><mrow><mn>10</mn></mrow><mrow><mo>−</mo><mn>4</mn></mrow></msup></mrow></semantics></math></inline-formula> (PBS) versus <inline-formula><math xmlns="http://www.w3.org/1998/Math/MathML" display="inline"><semantics><mrow><mn>8.6</mn><mo>×</mo><msup><mrow><mn>10</mn></mrow><mrow><mo>−</mo><mn>7</mn></mrow></msup></mrow></semantics></math></inline-formula> (AU). The three-order reduction in AU was attributed to the inhibitory effects inherent to complex biological fluids. These findings highlight the importance of the model-based quantitative analysis of EISCAP biosensors, enabling the accurate characterization of immobilized enzyme layers and guiding optimization for applications in realistic sample matrices.
Topik & Kata Kunci
Penulis (5)
Karen Simonyan
Astghik Tsokolakyan
Vahe Buniatyan
Artem Badasyan
Mkrtich Yeranosyan
Akses Cepat
- Tahun Terbit
- 2025
- Sumber Database
- DOAJ
- DOI
- 10.3390/s25216596
- Akses
- Open Access ✓