Safety Parameters of Quantum Molecular Resonance Devices During Thyroid Surgery: Porcine Model Using Continuous Neuromonitoring
Abstrak
ObjectivesQuantum molecular resonance (QMR) devices have been applied as energy-based devices in many head and neck surgeries; however, research on their use in thyroid surgery is lacking. This study aimed to investigate the safety parameters of QMR devices during thyroidectomy when dissection was adjacent to the recurrent laryngeal nerve (RLN).MethodsThis study included eight piglets with 16 RLNs, and real-time electromyography (EMG) signals were obtained from continuous intraoperative neuromonitoring (C-IONM). QMR bipolar scissor (BS) and monopolar unit (MU) were tested for safety parameters. In the activation study, QMR devices were activated at varying distances from the RLN. In the cooling study, QMR devices were cooled for varying time intervals, with or without muscle touch maneuver (MTM) before contacting with the RLN.ResultsIn the activation study, no adverse EMG change occurred when QMR BS and MU were activated at distances of 2 mm or longer from the RLNs. In the cooling study, no adverse EMG change occurred when QMR BS and MU were cooled in 2-second intervals or immediately after MTM.ConclusionQMR devices should be carefully used when performing RLN dissection during thyroid surgery. According to the activation and cooling safety parameters in this study, surgeons can avoid RLN injury by following standard procedures when using QMR devices.
Topik & Kata Kunci
Penulis (15)
Hsin-Yi Tseng
Tzu-Yen Huang
Yi-Chu Lin
Jia Joanna Wang
Jia Joanna Wang
How-Yun Ko
Cheng-Hsun Chuang
I-Cheng Lu
Pi-Ying Chang
Gregory W. Randolph
Gianlorenzo Dionigi
Gianlorenzo Dionigi
Ning-Chia Chang
Ning-Chia Chang
Che-Wei Wu
Akses Cepat
- Tahun Terbit
- 2022
- Sumber Database
- DOAJ
- DOI
- 10.3389/fendo.2022.924731
- Akses
- Open Access ✓