The Presence of Blood in a Strain Gauge Pressure Transducer Has a Clinical Effect on the Accuracy of Intracranial Pressure Readings

Emerson B. Nairon, BSA Jeslin Joseph, BS Abdulkadir Kamal, BSN, RN David R. Busch, PhD DaiWai M. Olson, PhD, RN, FNCS

Abstrak

IMPORTANCE:. Patients admitted with cerebral hemorrhage or cerebral edema often undergo external ventricular drain (EVD) placement to monitor and manage intracranial pressure (ICP). A strain gauge transducer accompanies the EVD to convert a pressure signal to an electrical waveform and assign a numeric value to the ICP. OBJECTIVES:. This study explored ICP accuracy in the presence of blood and other viscous fluid contaminates in the transducer. DESIGN:. Preclinical comparative design study. SETTING:. Laboratory setting using two Natus EVDs, two strain gauge transducers, and a sealed pressure chamber. PARTICIPANTS:. No human subjects or animal models were used. INTERVENTIONS:. A control transducer primed with saline was compared with an investigational transducer primed with blood or with saline/glycerol mixtures in mass:mass ratios of 25%, 50%, 75%, and 100% glycerol. Volume in a sealed chamber was manipulated to reflect changes in ICP to explore the impact of contaminates on pressure measurement. MEASUREMENTS AND MAIN RESULTS:. From 90 paired observations, ICP readings were statistically significantly different between the control (saline) and experimental (glycerol or blood) transducers. The time to a stable pressure reading was significantly different for saline vs. 25% glycerol (< 0.0005), 50% glycerol (< 0.005), 75% glycerol (< 0.0001), 100% glycerol (< 0.0005), and blood (< 0.0005). A difference in resting stable pressure was observed for saline vs. blood primed transducers (0.041). CONCLUSIONS AND RELEVANCE:. There are statistically significant and clinically relevant differences in time to a stable pressure reading when contaminates are introduced into a closed drainage system. Changing a transducer based on the presence of blood contaminate should be considered to improve accuracy but must be weighed against the risk of introducing infection.

Penulis (5)

E

Emerson B. Nairon, BSA

J

Jeslin Joseph, BS

A

Abdulkadir Kamal, BSN, RN

D

David R. Busch, PhD

D

DaiWai M. Olson, PhD, RN, FNCS

Format Sitasi

BSA, E.B.N., BS, J.J., RN, A.K.B., PhD, D.R.B., FNCS, D.M.O.P.R. (2024). The Presence of Blood in a Strain Gauge Pressure Transducer Has a Clinical Effect on the Accuracy of Intracranial Pressure Readings. https://doi.org/10.1097/CCE.0000000000001089

Akses Cepat

PDF tidak tersedia langsung

Cek di sumber asli →
Lihat di Sumber doi.org/10.1097/CCE.0000000000001089
Informasi Jurnal
Tahun Terbit
2024
Sumber Database
DOAJ
DOI
10.1097/CCE.0000000000001089
Akses
Open Access ✓