Development of a dose estimation code for BNCT with GPU accelerated Monte Carlo and collapsed cone Convolution method

Chang-Min Lee Hee-Seock Lee

Abstrak

A new method of dose calculation algorithm, called GPU-accelerated Monte Carlo and collapsed cone Convolution (GMCC) was developed to improve the calculation speed of BNCT treatment planning system. The GPU-accelerated Monte Carlo routine in GMCC is used to simulate the neutron transport over whole energy range and the Collapsed Cone Convolution method is to calculate the gamma dose. Other dose components due to alpha particles and protons, are calculated using the calculated neutron flux and reaction data. The mathematical principle and the algorithm architecture are introduced. The accuracy and performance of the GMCC were verified by comparing with the FLUKA results. A water phantom and a head CT voxel model were simulated. The neutron flux and the absorbed dose obtained by the GMCC were consistent well with the FLUKA results. In the case of head CT voxel model, the mean absolute percentage error for the neutron flux and the absorbed dose were 3.98% and 3.91%, respectively. The calculation speed of the absorbed dose by the GMCC was 56 times faster than the FLUKA code. It was verified that the GMCC could be a good candidate tool instead of the Monte Carlo method in the BNCT dose calculations.

Penulis (2)

C

Chang-Min Lee

H

Hee-Seock Lee

Format Sitasi

Lee, C., Lee, H. (2022). Development of a dose estimation code for BNCT with GPU accelerated Monte Carlo and collapsed cone Convolution method. https://doi.org/10.1016/j.net.2021.11.010

Akses Cepat

PDF tidak tersedia langsung

Cek di sumber asli →
Lihat di Sumber doi.org/10.1016/j.net.2021.11.010
Informasi Jurnal
Tahun Terbit
2022
Sumber Database
DOAJ
DOI
10.1016/j.net.2021.11.010
Akses
Open Access ✓