Compensation of synchronization error effect in testing machine calibration
Miha Hiti
Abstrak
The paper presents an evaluation of synchronization issues when performing material testing machine calibration using continuous loading. A procedure for the compensation of a rangewide error in calibrated force values is proposed based on a correction factor determined from static and transient errors difference at a chosen force step. The procedure was applied for a calibration of a testing machine with a 20 kN force transducer and 200 N/s to 2000 N/s loading rates. The results show good performance of the compensation technique, reducing calibration errors to below 1% in non-synchronized signals even under unknown filter and delay setting.