CrossRef Open Access 2021 2 sitasi

Improvement of the minimum detectable activity of a free release monitor for small articles

Serhii Pohuliai Igors Krainukovs

Abstrak

This paper presents results of the development of a small-sized free release monitor designed for the release of materials, various hand tools, equipment and instruments of nuclear enterprises and laboratories staff that weight up to 50 kg, from radiation control. To increase the registration sensitivity of controlled radionuclides, 12 scintillation units based on a 3"x3" sized NaI (Tl) crystal were used as gamma-radiation detectors. Volume of the measuring chamber of the monitor amounted to 200 L, the chosen thickness of the low-back ground shielding was 50 mm. The values of the minimum detect able activity of the designed monitor for the point sources 123I, 131I, 99mTc, 18F were better than 100 Bq with measurement time not exceeding 60 seconds.

Penulis (2)

S

Serhii Pohuliai

I

Igors Krainukovs

Format Sitasi

Pohuliai, S., Krainukovs, I. (2021). Improvement of the minimum detectable activity of a free release monitor for small articles. https://doi.org/10.2298/ntrp210222018p

Akses Cepat

PDF tidak tersedia langsung

Cek di sumber asli →
Lihat di Sumber doi.org/10.2298/ntrp210222018p
Informasi Jurnal
Tahun Terbit
2021
Bahasa
en
Total Sitasi
Sumber Database
CrossRef
DOI
10.2298/ntrp210222018p
Akses
Open Access ✓